If this message appears, disconnect
the charger and the device switches
off automatically. Please contact the
hotline, noting this error code.
If, for any reason, you experience
any other trouble with your device,
please contact the hotline mentioned
and approved by Compex Médical SA.
11a The small battery symbol has only two re-
ference marks: The batteries are losing their
charge. Interrupt the session and recharge the
device (see the following section).
11b The , symbol normally displayed
above the “+”/“–” key of channel 4 has disappea-
red ( ) and the small battery flashes: The
batteries are completely flat. It is no longer pos-
sible to use the device. Recharge it immediately
(see the following section).
Recharging
To recharge the Compex, first discon-
nect the electrode cables from the
device, then plug the charger in a
socket and lastly connect the stimula-
tor to the charger (see in section I:
“Connecting the charger”). The
charge menu illustrated below appears
automatically.
START
13a
14a
Fig.13
Fig.14
Fig.15
12a The Compex has been charging for 3 mi-
nutes and 8 seconds (a full charge may take from
1h30 to 2 hours with the rapid charger delivered
with your device).
12b Charging is in progress and the small
battery symbol is not completely full.
When charging is completed, total charge dura-
tion flashes and the small battery is completely
full. The Compex switches off automatically as
soon as you disconnect the charger.
8. Problems and solutions
Electrode fault
13a-14a The Compex “bleeps” and alternati-
vely displays the symbol of a couple of elec-
trodes and an arrow pointing to the channel
where a problem has been detected. Below, the
stimulator has detected an electrode fault on
channel 2
This message may mean that:
12a
12b
Fig.12
1) there is no electrode connected to this
channel;
2) the electrodes are old, worn out and/or the
contact is poor: try using new electrodes;
3) the electrode cable is defective: try it using
another channel. If cable shows a default again,
please replace.
Recharging error
A recharging error may happen du-
ring charging. The Compex “bleeps”
continuously and the following
screen is displayed: