Neurostimulation
All electrical specifications are given
for a charge between 500 and 1000
ohms per channel.
Channels: four independent and individually
adjustable channels electrically insulated from
each other and from ground
Impulse shape: constant rectangular current
compensated to eliminate any direct current
component to avoid any residual skin polarisation
Maximum current of an impulse: 120 milliamperes
Intensity increase interval:
– Manual stimulation intensity adjustment: 0-
999 (energy units)
– Impulse duration: 0.5 mA
From 50 to 400 microseconds
Maximum quantity of electricity per impulse: 96
microcoulombs (2 x 48 µC compensated)
Typical rise time for an impulse
3 microseconds (between 20% and 80% of
maximum current)
Impulse frequency
from 1 to 150 Hertz
8. Standards
The Compex is based directly on medi-
cal technology.
To guarantee your safety, the design,
manufacturing and distribution of Com-
pex are in conformity with the require-
ments of the European Directive
93/42/CEE.
The device is in conformity with the
standard for general safety rules for
electromedical devices IEC 60601-1, the
standard for electromagnetic compatibi-
lity IEC 60601-1-2 and the standard for
special safety rules for nerve and mus-
cle stimulators IEC 60601-2-10.
Current international standards (IEC
60601-2-10 AM1 2001) require that a
warning be given concerning the appli-
cation of electrodes to the thorax
(increased risk of cardiac fibrillation).
Directive 2002/96/CEE Waste electrical
and electronic equipment (WEEE).
9. Patent
Snap electrode: patent pending.
10. Normalised symbols
Important: Under some condi-
tions, the effective figure for
the stimulation pulses can exceed 10
mA and 10 V. The information given
in this handbook must be strictly ob-
served.
The Compex is a class II device
unit with its own internal electric
power, with type BF applied sections
Waste electrical and electronic
equipment (WEEE)
The “On/Off” switch is a multi-
function key:
Functions Symbol N°(according to CEI 878)
On/Off (2 positions, stable) 01-03
Waiting or preparation
for part of the unit 01-06
Stop (switching off) 01-10
11. Technical information
General points
Power supply NIMH rechargeable
battery (4,8 V ª 1200 mA/h)
Chargers: The only chargers used for
recharging the battery bear the follo-
wing information:
Type TR503-02-E-133A03
Input 90-264 VAC / 47-63Hz / 0.5A max.
Output 9V / 400mA / 6W
Type TR1509-06-E-133A03
Input 90-264 VAC / 47-63Hz / 0.5A max.
Output 9V / 1.4A / 15W
Type TR503-02-A-133A03
Input 90-264 VAC /47-63Hz / 0.5A max.
Output 9V / 400mA / 6W
Type TR1509-06-A-133A03
Input 90-264 VAC / 47-63Hz / 0.5A max.
Output 9V / 1.4A / 15W
Type TR503-02-U-133A03
Input 90-264 VAC / 47-63Hz / 0.5A max.
Output 9V / 400mA / 6W
Type TR1509-06-U-133A03
Input 90-264 VAC / 47-63Hz / 0.5A max.
Output 9V / 1.4A / 15W